Equally important, but
often overlooked, is the requirement to adjust the pH of samples prior
to dilution. This step is important because the test must be conducted
at a pH between 6.5 and 7.5 in order for the results to be accurate. In
some municipal labs, adjusting the pH of “real” samples
is never required as they always fall in the required range. Over time,
some of these labs even begin to stop checking the pH of their samples
altogether. While this may not cause a problem when testing plant effluents,
it can be a significant problem when analyzing a DMR-QA sample. In order
to prepare a homogenous, stable proficiency testing sample, ERA preserves
the Demand concentrate with acid. Once you have made the initial dilution
into deionized water, the sample pH will still be slightly acidic. If you
do not adjust the pH of this sample prior to preparing your analytical
dilutions, you run a high risk of failing the sample. The best way to adjust
the pH is to use a dilute solution of sodium hydroxide (~0.2 molar) and
add it one drop at a time to the diluted sample, mixing and then checking
the pH after each addition until the pH is between 6.5 and 7.5.
Here are a few additional keys to improving your BOD analyses:
* Make sure that your incubator temperature
remains consistent at 20±1°C.
The use of an independent thermometer is highly recommended.
* When analyzing CBOD be cautious about using plant influent as your seed
if it contains a high percentage of nitrifying bacteria. The nitrification
inhibitor you add to the sample will suppress the nitrifying bacteria and
you may not be left with enough carbonaceous bacteria to generate proper
oxygen depletion.
pH
While pH is generally considered a simple analysis,
it too requires some attention to detail to ensure consistently accurate
results. Don’t
let the test’s seeming simplicity make you take the importance of
these details for granted.
It is always better to run calibration, check and field samples at room
temperature rather than relying on automatic temperature compensation (ATC).
If you must analyze samples of different temperatures, make sure that your
meter has ATC and that it is working appropriately.
Use of fresh calibration buffers is recommended. You should never stick
your probe into your bottle of buffer or return buffer to the bottle after
it has been used once. By following the storage recommendations given by
the supplier and making sure to not contaminate the solution, you should
be able to use the buffer up until its expiration date. Using an independent
quality control sample will help you identify bad buffer solutions or a
probe or meter that is not operating correctly. Finally, you should always
bracket the concentration of your real samples with your calibration buffers.
By screening unknown samples before analysis you can identify which buffers
to use.
Total Residual Chlorine
Residual chlorine analysis can easily be compromised due to contamination
from chlorine present in tap water as well as the volatility of the analyte.
A few simple steps can minimize these and other sources of error in this
test.
* Make sure all glassware is cleaned and rinsed at least three times
with deionized water prior to use.
* Analyze a blank to ensure no contamination from water or glassware is
present.
* Samples should be analyzed as soon as possible. Proficiency testing (PT)
and quality control (QC) samples should be freshly diluted just before
they are to be analyzed.
* Make sure you are using fresh reagents. Reagent packets (powder pillows)
generally work very well, but you must make sure that the concentration
range of the powder pillow is appropriate for the sample you are testing.
* When diluting PT or QC samples, some glass pipets may not fit into the
neck of the glass ampule. A good solution is to use a clean, gas-tight
syringe capable of delivering 1 mL of concentrate.
Total Suspended Solids (TSS)
TSS is a pretty straight-forward test which many laboratories perform.
Here are a few tips to help ensure this analysis is in control.
* Filter an amount of sample to ensure that the resulting residue on
your filter is large enough to be weighed accurately yet small enough to
dry efficiently. A range of 10-200 mg of residue is recommended.
* Always use a 4-place analytical balance when weighing your filters before
and after use.
* You should always filter a blank (deionized water) sample and perform
corrective action if the results are outside of your normal observations.
* Ensure that your pans and filter papers are completely dry before use.
They should be stored in a dessicator after they are dried and before they
are used.
* Dry and weigh each sample at least twice to ensure that the filters are
completely dry.
* Dried pans and filters should be allowed to cool in a dessicator prior
to weighing.
Oil and Grease
Oil and Grease is a gravimetric analysis with some similarities to TSS.
A few of the same tips and some unique steps apply to this analysis.
As with TSS, it is important to use a 4-place analytical balance for
all weighings. Unlike TSS, where you choose the sample size to filter,
the entire oil and grease sample should always be used to ensure that all
of the analyte is captured. You also need to rinse the bottle and cap three
times with solvent to ensure that none of the sample remains attached to
the container.
Drying and weighing at least twice should be performed for the oil and
grease analysis as well, so that you can make sure the sample is completely
dry. It is also necessary to ensure that no water ends up in your collection
pans as the dissolved solids present can result in a high bias.
Coliforms
Like BOD, there are many variables that go into an accurate analysis
of coliforms. Over the years, ERA has answered hundreds of questions about
this analysis with many of the questions focusing on some of the same topics.
Positive Controls - For each new batch of media and method apparatus,
test a positive control to demonstrate that the media produces the expected
reaction to the organism under test. Reference cultures used for positive
controls should come from a national collection, organization, or manufacturer
recognized by an accrediting authority. These cultures may be a single-use
preparation or cultures maintained by procedures that ensure the continued
purity and viability of the organism.
Negative Controls - For each new batch of media and method apparatus,
test a negative control to demonstrate that the media does not demonstrate
a false-positive reaction.
Reference cultures used for negative controls should come from a national
collection, organization, or manufacturer recognized by an accrediting
authority. These cultures may be a single-use preparation or cultures maintained
by procedures that ensure the continued purity and viability of the organism.
Method Evaluation - All methods in use in the laboratory should be evaluated
for their ability to produce acceptable results prior to first use. For
quantitative microbiology methods this can be accomplished by participation
in an approved Proficiency Testing program or the analysis of a Quality
Control sample.
Hydration of Pellet - The initial hydration of the gelatin tablet supplied
with ERA samples should be performed following the instructions provided.
Please do not deviate from the instructions as the certified values and
acceptance intervals are based on customer data and deviations could result
in an incorrect result.
Once hydration of the pellet is complete, perform your analysis immediately.
Waiting more than 30 minutes to perform your analysis could have an affect
on results. Gently shake the sample prior to taking aliquots for analysis.
Culture Media - Quality of culture media is critical.
Never prepare media from raw ingredients if a source of dehydrated media
is available. If preparing dehydrated media, follow instructions closely.
Always check pH and make any adjustments that are necessary. Always evaluate
media prior to first use. Never use media outside of its expiration date
and never use media that has not been stored according to the manufacturer’s
specifications. Confirm that prepared media and dehydrated media ingredients
and proportions match the specifications for your method.
Dilution and Rinse Water - Only hydrate the gelatin tablet using the
phosphate buffer supplied by ERA. To perform dilutions and apparatus rinsing,
do not use DI or distilled water. ERA recommends Phosphate Buffer or Peptone
water as it is not uncommon for DI or distilled water to cause inhibitory
effects. See Standard Methods 9050C 20th Edition for the preparation of
Buffered water or Peptone water. |